Author :
Publisher :
Page : 84 pages
File Size : 42,18 MB
Release : 1979
Category : Aerosols
ISBN :
Studies were conducted on the use of Nuclepore filters to perform a dichotomous particle size separation in ambient aerosols. Optimization of the resulting stacked filter unit was undertaken, with incorporation of information from theoretical analyses, laboratory studies, and tests in a variety of ambient conditions. The stacked filter unit was improved by development of antibounce coatings for the coarse filter, incorporation of a new fine filter developed by the Nuclepore Corporation, and addition of a pneumatic constant flow device. The unit has been shown to quantitatively collect particles between 2.5 micrometers and 15 micrometers on the coarse filter, and 0 to 2.5 micrometers on the fine filter, with a separation curve similar to that of the human respiratory tract. Stacked filter units weigh about 15 kg., and total cost of components for the basic unit is about 200 dollars.